Section 4
Options
Immediate Taper-Down: TPN
This option is available in the TPN delivery mode only. It allows you to stop
infusion early by immediately tapering down. If the pump is in LL0 or LL1,
the taper-down period can be changed from its originally programmed
value, even if it is 0. You can select a taper-down period that is equal to or
less than the time remaining in the infusion profile. If the pump is in LL2,
you can use the Immediate Taper-Down Option, but you cannot change the
taper-down period.
To access the Immediate Taper-Down option:
>
The pump must be running
>
The pump must be delivering at the plateau rate
>
There must be at least 10 minutes left in the infusion period
Prime
The Prime option is used in all delivery modes to pump fluid through the
fluid path to remove air bubbles prior to connecting to the patient. The
pump must be stopped and in LL0 or LL1 to prime.
Fluid delivered with the priming feature is subtracted from the reservoir
volume value, but is not added to the amount given (since this fluid is not
delivered to the patient). Priming is not allowed when the reservoir
volume value is 0.0 ml.
To prime, select Prime from the Options menu, then follow the steps on
page 42, beginning with step 2.
Time Remaining: Intermittent
This option is available in the Intermittent delivery mode only. It allows you
to view the time remaining in the current dose if one is being delivered and
the time remaining in the cycle. The pump may be running or stopped and
in any lock level. See page 34 for instructions.