Checking the system adjustment
Sirona Dental Systems GmbH
9.1
Test exposures
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
90
D 3352.031.03.18.02
4.
Evaluate the X-ray image:
– The shadow of the center needle, the needle
image and the auxiliary line must be coincident
and located behind each other.
A tolerance (offset of needle from the central
auxiliary line) of
± 0.75
mm is admissible A.
– Distance A1 must be 88.6
± 1
mm A.
– Distances A2 must be identical, each being 44.3
± 0.5
mm A.
– A white border surrounding the image on all
sides must be visible B.
IMPORTANT
If one of these criteria is not fulfilled
C
, the pan symmetry
must be adjusted. The procedure for adjusting the unit is de
-
scribed in the Appendix (see section 11.2) and in the OR
-
THOPHOS XG Service Manual.
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......