8
Startup for USA/Canada only
Sirona Dental Systems GmbH
8.4
kV – verification / Exposure Time Verification
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
80
D 3352.031.03.18.02
kV – verification
Analyzing measurements
➢
Read the voltage values on the Mult-O-Meter.
The value for the voltage displayed on the
Mult-O-Meter must correspond to the 62 kV
selected in the service routine. The permissible
tolerance is ± 10 %..
If the measured values
do not fall within the
permissible tolerance
range, replace the tube
assembly (see Service Manual).
If the measured values fall within the permissible
tolerance range, finalize the measurement.
Exposure Time Verification
Analyzing measurements
➢
Read the radiation time on the Mult-O-Meter.
The value for the radiation time displayed on the
Mult-O-Meter must correspond to the radiation
time of
0.5s
selected in the service routine. The
permissible tolerance is ± 10 %.
If the measured radiation time does
not
fall within
the permissible tolerance, replace the
tube
assembly
(see Service Manual).
If the measured radiation time falls within the
permissible tolerance, finalize the measurement.
Finalizing the measurement
➢
Exit the service routine.
➢
Switch off the unit (see Operating Instructions).
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......