7
Initial startup
Sirona Dental Systems GmbH
7.3
Switching the units ON
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
68
D 3352.031.03.18.02
WARNING
When performing the following test, be sure to observe
the radiation protection regulations applicable in your
country (see operating instructions).
It is prohibited for any person to be positioned in the
unit when it is switched on.
NOTICE
If the room height is less than 2.27 m (89 3/8“) (2.30 m (90
1/2“) with floor stand), you must limit the maximum travel
height of the unit (see section 11.1.7).
1.
Switch the unit
ON
.
– All LEDs and the LED display on the Multipad light
up briefly.
– The green LED at the top of the Multipad
➋
is lit
permanently as long as the unit is ON.
– The initialization status is visualized by a progress
indicator while the unit executes a self-adjustment
routine (approx. ½ min.). At the same time, the
rotating element rotates briefly clockwise and coun
-
terclockwise and the diaphragm is positioned. The
forehead and temple supports on the panoramic
unit open and close and then stop moving in fully
opened position.
– Once the self-adjustment routine is completed, help
message H301 prompts you to move the unit to the
starting position.
2.
Press the
R key
to move the unit back to the starting
position.
•
Switch the PC
ON
and start SIDEXIS XG.
7.3 Switching the units ON
1.
?
?
2.
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......