Sirona Dental Systems GmbH
11
Appendix
Installation Instructions ORTHOPHOS XG 3
11.3
Demo mode
60 51 416 D 3352
D 3352.031.03.18.02
137
11.3.1Switching the demo mode ON
When operated in demo mode, the unit must not release
any radiation. For this reason, you must take the following
safety measures:
1.
Remove the cover of the tube assembly.
2.
Loosen screws
A
and remove cover plate
B
.
3.
Set DIP switch
S2
(DX6) to
position 2
.
4.
Pull cable
L5
(X-RAY) off of connector
J6
(DX6).
Radiation release is now no longer possible.
5.
Check the mode using the Info menu.
Demo mode: ON means that: The demo mode is
switched ON
(Radiation release is not possible)
Demo mode: OFF means: The demo mode is switched
OFF
(Radiography, X-ray radiation are possible!)
•
Reattach cover plate
B
and the tube assembly covers
by following the dismantling procedure in reverse order.
11.3.2Switching the demo mode OFF
1.
Remove the cover of the tube assembly.
2.
Loosen screws
A
and remove cover plate
B
.
3.
Set DIP switch
S2
(DX6) to
position 1
.
4.
Connect cable
L5
(X-RAY) to connector
J6
(DX6).
Radiation release is now once again possible.
5.
Check the mode using the Info menu.
Demo mode: ON means that: The demo mode is
switched ON
(Radiation release is
not
possible)
Demo mode: OFF means: The demo mode is switched
OFF
(Radiography, X-ray radiation are possible!)
•
Reattach cover plate
B
and the tube assembly covers
by following the dismantling procedure in reverse order.
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......