Adjusting the panoramic X-ray unit
60 51 416 D 3352
D 3352.031.03.18.02
131
8.
If all criteria are fulfilled and the current image is identi
-
cal to the ideal image
A + B
, then save the values:
Click
SAVE
VALUES
IMPORTANT
In SIDEXIS versions V01.50 and higher, the values for
S1
,
S2
and
S3
in the
PAN
-
SYMMETRY
submenu are set
equal
to zero
with a correct adjustment, i.e. if they are within the
permissible tolerance. For versions < V01.50, the values in
the text boxes may deviate slightly from zero even with a
correct adjustment.
•
The adjustment of the panoramic X-ray unit is now
completed.
8.
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......