11
Appendix
Sirona Dental Systems GmbH
11.2
Adjusting the panoramic X-ray unit
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
126
D 3352.031.03.18.02
11.2.7Adjusting the PAN symmetry
•
Insert the needle phantom in the bite block holder of the
panoramic X-ray unit (see page 112).
1.
Go to the
PAN
-
SYMMETRY
submenu.
2.
To make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION
The exposure dialog box showing the exposure status
appears in Sidexis.
The initialization status is visualized by a progress indi
-
cator on the Multipad.
The initialization procedure is completed when the ex
-
posure parameters of service routine
S010.2
(60 kV / 3 mA; 14.1 s) are displayed and the progress
indicator disappears.
3.
Take an exposure (60 kV/3 mA):
– Press the
R key
on the Multipad to move the unit
back to the starting position.
– Press the release button. Hold down the release
button until image acquisition is completed and the
preview image appears in the exposure dialog box.
1.
2.
in SIDEXIS
on the Multipad
6 0 / 3
1 4 . 1
3.
T
R
Prog.
S
kV
mA
G
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......