11
Appendix
Sirona Dental Systems GmbH
11.2
Adjusting the panoramic X-ray unit
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
118
D 3352.031.03.18.02
7.
Confirm the displacement of the sensor:
Click
OK
The exposure dialog box showing the exposure status
appears in Sidexis.
The initialization status is visualized by a progress indi
-
cator on the Multipad.
The initialization procedure is completed when the ex
-
posure parameters of service routine
S010.1
(60 kV / 3 mA; 0.6 s) are displayed and the progress in
-
dicator disappears.
in SIDEXIS
7.
on the Multipad
6 0 / 3
0 . 6 0
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......