Sirona Dental Systems GmbH
11
Appendix
Installation Instructions ORTHOPHOS XG 3
11.2
Adjusting the panoramic X-ray unit
60 51 416 D 3352
D 3352.031.03.18.02
109
11.2.1Diaphragm/system adjustment menu
The
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
menu guides
you through the adjustment of the panoramic unit. This ser
-
vice routine is started from
SIDEXIS XG
:
U
TILITIES
C
ONSTANCY
TEST
XCXP
S
ELECT
X-
RAY
DEVICE
S
ERVICE
EXPOSURE
P
ASSWORD
INPUT
S
ELECT
X-
RAY
COMPONENT
D
IAPHRAGM
TEST
EXPOSURE
IMPORTANT
The
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
menu is pass
-
word-protected. As password, enter the first four digits of
the current system date (PC) in reverse order.
Example: On 05/30/2004, the service password is
5003
IMPORTANT
The
SELECT
X
-
RAY
DEVICE
and
SELECT
X
-
RAY
COMPO
-
NENT
dialogs appear only if several x-ray devices or x-ray
components are available for selection.
IMPORTANT
When you open the
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
menu, the unit switches from the user mode to the PC ser
-
vice mode logged by the PC. The service mode is indicated
by the
SERVICE
display on the Multipad (
2.
).
In the
PC service mode
, the control options that are avail
-
able on the Multipad are determined by SIDEXIS and the
currently selected service routine. General control of the
unit by means of the Multipad (as in the user mode) is not
possible in this mode.
The
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
menu has
5 submenus:
•
PAN - Sensor adjustment
•
PAN - Diaphragm
•
PAN - Symmetry
•
PAN - Reset adjustment
•
Sirona Service Sheet (for internal use only)
You can change between the individual submenus by click
-
ing the
menu tabs
with the mouse. To quit the
DIA
-
PHRAGM
/
SYSTEM
ADJUSTMENT
menu, click
CANCEL
.
11.2Adjusting the panoramic X-ray unit
1.
2.
S E R V I C E
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......