3
Installation: Panoramic X-ray unit
Sirona Dental Systems GmbH
3.4
Installing the floor stand (option 2)
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
30
D 3352.031.03.18.02
CAUTION
For installation with the floor stand, the unit remains lying on
the pallet until the floor stand has been completely assem
-
bled. Only then may the unit be installed. For enhanced rep
-
resentation, some of the following drawings are shown in
the standing state.
1.
Remove the surrounding packaging, the two lateral sty
-
rofoam parts and profile cover
A
.
NOTICE
The center styrofoam part should remain attached to the
unit for protection.
2.
Loosen the nuts
B
(with washers) on both sides of the
pallet (or the wooden support).
Take off the pallet (or
the wooden support). Remove the threaded bolt
C
.
IMPORTANT
The nuts
D
on the unit may remain inside the unit when the
threaded rods are removed. Remove the nuts.
3.
Remove the carrying handles
E
on the lower side of the
pallet from the holders
F
.
4.
Loosen screws
G
and remove the holders
F
.
3.4 Installing the floor stand (option 2)
1.
A
B
C
C
B
C
D
2.
4.
E
G
F
Must be dis
-
mantled ly
-
ing on its
3.
E
F
F
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......