Installation: Panoramic X-ray unit
Installation Instructions ORTHOPHOS XG 3
3.3
Wall mounting (standard/option 1)
60 51 416 D 3352
D 3352.031.03.18.02
27
•
Move the panoramic X-ray unit into its installation posi
-
tion at the wall. Hold the unit laterally at the styrofoam
packaging to do this.
NOTICE
SIRONA recommends leaving the styrofoam packaging on
the unit during the entire installation procedure!
If due to on-site conditions it is unavoidable to remove the
styrofoam packaging already at this point, you may move
the unit by
carefully
grasping the bite block bar and the
stand.
6.
Fasten the panoramic X-ray unit to the upper wall hold
-
er.
– Insert the screws
E
into the groove.
– Screw the panoramic X-ray unit firmly onto the wall
holder using the washers and nuts
F
.
IMPORTANT
The wall holder must be flush with the upper edge of the
unit.
7.
Insert screw
G
through washer
H
and then through
the wall holder and into the stand from the rear.
8.
Fit profile clamp
J
onto screw
G
from the other (front)
side and screw nut
K
onto the screw.
Tighten nut
K
firmly.
6.
E
F
2x M 8x30
2x
δια.
8.4
2x M 8
7.
1x M 8x50
1x
δια.
8.4
G
H
E
F
G
H
8.
D
1x M 8
J
Profile clamp
1x
K
K
J
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......