Sirona Dental Systems GmbH
3
Installation: Panoramic X-ray unit
Installation Instructions ORTHOPHOS XG 3
3.3
Wall mounting (standard/option 1)
60 51 416 D 3352
D 3352.031.03.18.02
25
CAUTION
If the setup site for the unit is carpeted, the carpeting must
be removed.
Wall plugs! Each wall plug must withstand an extraction
force of 700 N.
The wall construction must be suitable for installation of the
unit (see "On-site installation, dimensions, technical data")
In case of mounting on weight-bearing wooden con
-
structions:
Use the enclosed wood screws and washers from the
mounting kit for mounting the unit on weight-bearing wood
structures.
4.
Mount the upper wall holder.
2x 8x80
2x
δια.
8.4
4.
Δια.
3/8
″
Δια.
10 mm
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......