9
Checking the system adjustment
Sirona Dental Systems GmbH
9.1
Test exposures
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
86
D 3352.031.03.18.02
9.1.3 Checking the diaphragm adjustment
WARNING
When performing the following tests, be sure to ob
-
serve the radiation protection regulations applicable in
your country (see Operating Instructions).
•
Plug the sensor into the sensor slot on the panoramic
X-ray unit.
IMPORTANT
The needle phantom must
not
be fitted on the bite
block holder on the panoramic X-ray unit.
•
Open the
TEST
EXPOSURE
menu
(see section 9.1.1).
1.
Select the
PAN
-
DIAPHRAGM
submenu.
2.
To make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION
The exposure dialog box showing the exposure status
appears in Sidexis.
The initialization status is visualized by a progress indi
-
cator on the Multipad.
The initialization procedure is completed when the ex
-
posure parameters of service routine
S030.2
(60 kV / 3 mA; 0.20 s) are displayed and the progress
indicator disappears.
3.
Take an exposure (60 kV/3 mA):
– Press the
R key
on the Multipad to move the unit
back to the starting position.
– Press the release button. Hold down the release
button until image acquisition is completed and the
preview image appears in the exposure dialog box.
3.
T
R
Prog.
S
kV
mA
G
1.
2.
in SIDEXIS
on the Multipad
6 0 / 3
0 . 2 0
Summary of Contents for orthophos xg 3
Page 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Page 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Page 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Page 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Page 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Page 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Page 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Page 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Page 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Page 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Page 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Page 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Page 141: ......