OrbiSinusSpi
Indications:
Spiral mode for orbita and paranasal sinuses studies,
e.g. sinusitis, mucocele, pneumatization, polyposis,
tumor, corrections etc.
Two recon jobs are predefined for image reconstruction
of the soft tissue and bone structure.
Head
OrbiSinusSpi
2
nd
reconstr.
kV
120
Effective mAs
100
Slice collimation
0.75 mm
Slice width
4.0 mm
1.0 mm
Feed/Rotation
5.0 mm
Rotation time
0.75 sec.
Kernel
H30s
H60s
Increment
4.0 mm
0.7 mm
Image order
ca-cr
CTDI
w
22.4 mGy
Effective dose
Male: 0.8 mSv
Female: 0.8 mSv