48
NeonateBody
Indications:
Spiral mode for routine neonate body studies, e.g.
tumors, abnormalities, malformations, abscesses, etc.
NeonateBody
kV
80
Effective mAs
25
Slice collimation
1.5 mm
Slice width
6.0 mm
Feed/Rotation
24 mm
Rotation time
0.5 sec.
Kernel
B30f
Increment
6.0 mm
Image order
cr-ca
CTDI
w
0.55 mGy
Effective dose*
Male: 0.5 mSv
Female: 0.7 mSv
Tips
• You can modify the slice width for image reconstruc-
tion according to the clinical indications.
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/Access site
Total amount
1 – 2 ml per kg of body weight
* The conversion factor for an 8-week-old,
and a scan range of 150 mm was used.
Children