134
Spine
RoutineSeq
Indications:
Sequence mode for spine studies, e.g. prolapse,
degenerative changes, trauma, tumors etc.
RoutineSeq
kV
120
Effective mAs
200
Slice collimation
0.75 mm
Slice width
1.5 mm
Feed/Scan
9.0 mm
Rotation time
1.0 sec.
Kernel
B30s
Cycle time
3.0 sec.
Image order
cr-ca
CTDI
w
16.8 mGy
Effective dose
Male: 0.4 – 0.5 mSv
Female: 0.7 – 1.5 mSv