InnerEarSpi
Indications:
Spiral mode for inner ear Ultra High-Resolution studies
using 0.75 mm slice collimation.
E.g. Inflammatory changes, tumorous processes
of pyramids, cerebellopontine angle tumors, post-
traumatic changes, etc.
Note: UHR mode requires a 250 mm scan FOV.
And it is mandatory to position the patient in the
center of the scan Field of View.
Head
InnerEarSpi
kV
120
Effective mAs
120
Slice collimation
0.75 mm
Slice width
0.75 mm
Feed/Rotation
5.0 mm
Rotation time
1.0 sec.
Kernel
U90u
Increment
0.5 mm
Image order
ca-cr
CTDI
w
26.9 mGy
Effective dose
Male: 0.4 mSv
Female: 0.4 mSv
The images can be loaded in the 3D Card, MPR. Select
the Range function and a predefined range, linked to
the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.