140
AngioRunOff
Indications:
For CTA studies
A range of 100 cm will be done in 21.4 s.
AngioRunOff
2
nd
reconstr.
kV
120
Effective mAs
130
Slice collimation
1.5 mm
Slice width
6.0 mm
2.0 mm
Feed/Rotation
24.0 mm
Rotation time
0.5 sec.
Kernel
B30s
B20s
Increment
6.0 mm
1.5 mm
Image order
cr-ca
cr-ca
CTDI
w
9.1 mGy
Effective dose
Male: 4.4 mSv
Female: 3.2 mSv
Extremities
Summary of Contents for SOMATOM Sensation 16
Page 1: ...SOMATOM Sensation 16 Application Guide Routine Protocols Software Version A50 ...
Page 7: ...7 Content ...
Page 23: ...23 General ...
Page 29: ...29 Children ...
Page 49: ...49 Children ...
Page 55: ...55 Children ...
Page 94: ...94 Shoulder Overview ShoulderSpi Spiral mode for HiRes bone studies and soft tissues ...
Page 115: ...115 Abdomen ...