64
ExtremitySpiral
Indications:
Spiral mode for extremity studies, e.g. post-traumatic
changes, tumors, malformations, etc.
ExtremitySpiral 2
nd
reconstr.
kV
120
Effective mAs
*
Slice collimation
0.75 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
12.0 mm
Rotation time
0.75 sec.
Kernel
B30s
B60s
Increment
5.0 mm
0.7 mm
Image order
cr-ca
* The mAs should be adjusted to the body weight
of the child.
Children
Body weight
kV
mAs
CTDI
w
(mGy)
< 15 kg
120
21
1.7
15 – 34 kg
120
35
2.9
35 – 54 kg
120
70
5.9