135
Spine
ObeseSeq
Indications:
Sequence mode for spine studies of the obese
patients when higher dose are needed, e.g. prolapse,
degenerative changes, trauma, tumors etc.
ObeseSeq
kV
120
Effective mAs
400
Slice collimation
0.75 mm
Slice width
1.5 mm
Feed/Scan
9.0 mm
Rotation time
1.5 sec.
Kernel
B30s
Cycle time
3.0 sec.
Image order
cr-ca
CTDI
w
33.6 mGy
Effective dose
Male: 0.7 – 0.9 mSv
Female: 1.4 – 2.9 mSv