AbdMultiPhase
Indications:
Combination of CTA and soft tissue studies. E. g.
a range of 20 cm including liver, pancreas and kidney:
arterial phase acquired with 0.75 mm slice collimation
in 8.9 s, image reconstructed in 1.0 mm for post-
processing, and 5 mm for soft tissue enhancement.
Same range followed by a 1.5 mm slice collimation
acquisition in 4.8 s for venous phase, image recon-
structed in 5 mm for soft tissue only.
120
Abdomen
Native
Arterial
Venous
phase
phase
kV
120
120
120
Effective mAs
120
140
120
Slice collimation 1.5 mm
0.75 mm
1.5 mm
Slice width
5.0 mm
5.0 mm/
5.0 mm
1.0 mm
Feed/Rotation
24.0 mm
12.0 mm
24.0 mm
Rotation time
0.5 sec.
0.5 sec.
0.5 sec.
Kernel
B30f
B30f
B30f
Increment
5.0 mm
5.0 mm/
5.0 mm
0.7 mm
Image order
cr-ca
cr-ca
cr-ca
CTDI
w
8.4 mGy
10.9mGy
8.4 mGy
Effective dose
Male:
Male:
Male:
3.6 mSv
6.3 mSv
4.8 mSv
Female:
Female:
Female:
7.5 mSv
9.7 mSv
7.5 mSv