
Head
ThinSliceSeq
RoutineSeq
kV
120
120
Effective mAs
260
260
Slice collimation 0.75 mm
1.5 mm
Slice width
4.5 mm
9.0 mm
Feed/Scan
9.0 mm
18.0 mm
Rotation time
1.0 sec.
1.0 sec.
Kernel
H40s
H40s
Image order
ca-cr
ca-cr
CTDI
w
58.2 mGy
60.3 mGy
Effective dose
Male: 0.7 mSv
Male: 1.6 mSv
Female: 0.8 mSv Female: 1.8 mSv
PosteriorFossaOptimization Filter
PFO: To reduce beam-hardening artifacts in head
images, particularly in the base of the skull, use the
Posterior Fossa Optimization (PFO) filter.
If you want to set up the scan with only one range,
i. e. either 12 x 0.75 mm or 12 x 1.5 mm, you can
easily delete one mode by clicking the chronicle with
the right mouse button, and select cut.
Contrast medium IV injection
Start delay
60 sec.
Flow rate
2 ml/sec.
Total amount
50 – 60 ml
77
Summary of Contents for SOMATOM Sensation 16
Page 1: ...SOMATOM Sensation 16 Application Guide Routine Protocols Software Version A50 ...
Page 7: ...7 Content ...
Page 23: ...23 General ...
Page 29: ...29 Children ...
Page 49: ...49 Children ...
Page 55: ...55 Children ...
Page 94: ...94 Shoulder Overview ShoulderSpi Spiral mode for HiRes bone studies and soft tissues ...
Page 115: ...115 Abdomen ...