117
Abdomen
AbdRoutine
kV
120
Effective mAs
140
Slice collimation
1.5 mm
Slice width
5.0 mm
Feed/Rotation
24.0 mm
Rotation time
0.5 sec.
Kernel
B30f
Increment
5.0 mm
Image order
cr-ca
CTDI
w
10.0 mGy
Effective dose
Male: 5.8 mSv
Female: 8.9 mSv
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E.g. when both non-contrast and contrast studies are
required.
Contrast medium IV injection
Start delay
50 – 60 sec.
Flow rate
4.0 ml/sec.
Total amount
100 ml
Summary of Contents for SOMATOM Sensation 16
Page 1: ...SOMATOM Sensation 16 Application Guide Routine Protocols Software Version A50 ...
Page 7: ...7 Content ...
Page 23: ...23 General ...
Page 29: ...29 Children ...
Page 49: ...49 Children ...
Page 55: ...55 Children ...
Page 94: ...94 Shoulder Overview ShoulderSpi Spiral mode for HiRes bone studies and soft tissues ...
Page 115: ...115 Abdomen ...