131
Pelvis
PelvisRoutine
kV
120
Effective mAs
140
Slice collimation
1.5 mm
Slice width
5.0 mm
Feed/Rotation
24.0 mm
Rotation time
0.5 sec.
Kernel
B30f
Increment
5.0 mm
Image order
cr-ca
CTDI
w
9.8 mGy
Effective dose
Male: 3.3 mSv
Female: 4.8 mSv
* If the examination performed requires a full urinary
bladder, wait at least 3 minutes following IV admini-
stration of the contrast medium.
Contrast medium IV injection
Start delay
50 sec.*
Flow rate
2.0 – 3.0 ml/sec.
Total amount
100 – 120 ml
Summary of Contents for SOMATOM Sensation 16
Page 1: ...SOMATOM Sensation 16 Application Guide Routine Protocols Software Version A50 ...
Page 7: ...7 Content ...
Page 23: ...23 General ...
Page 29: ...29 Children ...
Page 49: ...49 Children ...
Page 55: ...55 Children ...
Page 94: ...94 Shoulder Overview ShoulderSpi Spiral mode for HiRes bone studies and soft tissues ...
Page 115: ...115 Abdomen ...