PulmonaryEmboli 2
nd
reconstr.
kV
120
Effective mAs
100
Slice collimation 0.75 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
15.0 mm
Rotation time
0.5 sec.
Kernel
B30f
B80f
Increment
5.0 mm
0.7 mm
Image order
cr-ca
CTDI
w
7.8 mGy
Effective dose
Male: 3.0 mSv
Female: 3.9 mSv
103
* CARE Bolus may be used to optimize the bolus
timing.
Set the ROI for monitoring scan in the truncus
pulmonalis with triggering threshold of 120 HU,
or use manual triggering.
Contrast medium IV injection
Start delay
12 – 15 sec.*
Total amount
100 – 120 ml
Thorax