* CARE Bolus may be used to optimize the bolus
timing. Set the ROI for monitoring scan in the abdo-
minal aorta at the beginning of the scan range
with triggering threshold of 120 HU, or use manual
triggering.
If Topo length 1024 mm is not long enough, you can
also choose the 1540 mm long Topogram.
Position the patient as feet first. Bend the feet together
if necessary.
The data acquired can also be used for image recon-
struction of soft tissue, done by the first reconstruction.
If the Pelvis region is included in the scan range, we
recommend at least 120 kV.
MIPthin: AngioRunOff
coronal
Image thickness
3
Distance between images
3
Number of images
50
For the 2
nd
reconstruction the Autoload into the
MIPthin Range on the 3D Card is activated. The images
will be automatically loaded into 3D, MIPthin, and
coronal MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Extremities
Contrast medium IV injection
Start delay
25 – 30 sec.*
Flow rate
3.0 – 3.5 ml/sec.
Total amount
150 ml