44
NeckRoutine
Indications:
Spiral mode for routine neck studies, e.g. tumors,
lymphoma, abscesses, etc.
NeckRoutine
2
nd
reconstr.
kV
120
Effective mAs
*
Slice collimation
1.5 mm
Slice width
6.0 mm
3.0 mm
Feed/Rotation
18 mm
Rotation time
0.75 sec.
Kernel
B30s
B60s
Increment
6.0 mm
3.0 mm
Image order
cr-ca
* The mAs should be adjusted to the age of the child.
* The conversion factor for an 8-week-old,
and a scan range of 80 mm was used.
** The conversion factor for a 7-year-old and a scan
range of 120 mm was used.
Age
kV
mAs
CTDI
w
Effective dose
(mGy)
(mSv)
< 3 years
120 40
2.9
Male: 0.7*
Female: 0.8*
3 years –
120 60
4.3
6 years
6 years –
120 90
6.5
Male: 0.8**
12 years
Female: 0.8**