
143
CombiSpi
2
nd
reconstr.
kV
120
Effective mAs
90
Slice collimation
0.75 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
13.5 mm
Rotation time
0.75 sec.
Kernel
B30s
B60s
Increment
5.0 mm
0.7 mm
Image order
cr-ca
CTDI
w
7.6 mGy
Effective dose
Male: 0.01 mSv
Female: 0.01 mSv
MPRthick: CombiSpi
coronal
Image thickness
2
Distance between images
2
Number of images
30
The 2
nd
reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined range,
linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Extremities
Summary of Contents for SOMATOM Sensation 16
Page 1: ...SOMATOM Sensation 16 Application Guide Routine Protocols Software Version A50 ...
Page 7: ...7 Content ...
Page 23: ...23 General ...
Page 29: ...29 Children ...
Page 49: ...49 Children ...
Page 55: ...55 Children ...
Page 94: ...94 Shoulder Overview ShoulderSpi Spiral mode for HiRes bone studies and soft tissues ...
Page 115: ...115 Abdomen ...