60
SpineSpiral
Indications:
Spiral mode for spine studies, e.g. post-traumatic
changes, tumors, malformations, orthopedic indi-
cations, etc.
SpinSpiral
2
nd
reconstr.
kV
120
Effective mAs
*
Slice collimation
1.5 mm
Slice width
3.0 mm
3.0 mm
Feed/Rotation
18.0 mm
Rotation time
0.75 sec.
Kernel
B30s
B60s
Increment
3.0 mm
3.0 mm
Image order
cr-ca
* The mAs should be adjusted to the body weight
of the child.
Children
* The conversion factor for an 8-week-old,
and a scan range of 100 mm was used.
** The conversion factor for a 7-year-old and a scan
range of 200 mm was used.
Body weight
kV
mAs
CTDI
w
Effective dose
(mGy)
(mSv)
< 15 kg
120 26
1.9
Male: 1.0*
Female: 1.0*
15 – 24 kg
120 55
4.0
25 – 34 kg
120 90
6.5
Male: 3.2**
Female: 6.4**
35 – 44 kg
120 130
9.4
45 – 54 kg
120 200
14.4