Page 30
SITRANS LR 300 – INSTRUCTION MANUAL
7ML19985CL02
mmm
mm
Inst
allat
ion
Mounting: Horn Antennas or Shielded Rod
The end of the horn, or the end of the shield section, should protrude a minimum of
10mm (0.5”) to avoid interference with the nozzle.
Mounting: Waveguide Antenna
This option is recommended for products with
ε
r
lower than 3. See P655 on page 68 for
the related propagation factor.
Notes:
• You can connect a maximum of two waveguides together.
• This option is recommended only for clean liquids, and only on vessels without an
agitator, with no turbulence.
• Horizontal stress on this antenna must be avoided, otherwise mechanical support
may be required.
• Process temperature and pressure capabilities are dependent upon information on
the process device tag. The reference drawing listed on the tag is available upon
request.
10 mm (0.5")
10 mm (0.5")
Vent hole
minimum 80mm (3")