Appendix VII
Parameter Structure
The diagram below shows a level application and the parameters involved.
Sensor Value
The value produced by the echo processing, which represents the distance from the
Sensor Reference Point to the target.
Sensor Unit
Units for Sensor Value, calibration points and other sensor-related parameters.
Level, Primary Value
The level as a percentage of Span (P007): (units are %).
Level, Primary Value Unit
Always %.
Sensor Reference Point
The point to which all of the above parameters are referenced, which is the flange face.
1.
Range extension can be set beyond the tank bottom.
P927
Distance
P922
Space
P921 Level /
P924 Volume
P0
07
Spa
n
P006
Empty
P801 Range Extension
1
Sensor
Reference
Point
Se
ns
or
V
al
ue
Profibus Profile
(arrows indicate possible settings)
100%
0%
Le
ve
l / P
rim
ar
y V
alu
e
Traditional Parameters
(individual parameters are named)
SITRANS LR 300.book Page 121 Monday, June 10, 2002 7:18 PM