Appendix VII
Example 2:
Status Word
Values in hex notation
Description
0x80
Data is GOOD.
0x4C
Initial value when the unit is first started up, before a valid reading is
taken.
1
1.
At this time the LCD screen display shows ----.
0x44
The data displayed is uncertain: it is the last usable data recorded before
an LOE state occurred.
0x10
The Failsafe timer has expired: this could be caused by LOE or by a sensor
malfunction: value is BAD.
0x04
There is an error in the configuration of the function blocks in Profibus PA
2
.
2.
This could happen when a firmware download has been done, but a system reset
(P999) has not been done. This could also happen if the function blocks are not con-
figured properly using PDM or acyclic services.
0X1E
The function block has been placed out of service. (You will see this only if
you read status word via acyclic services, after placing the function block
out of service
3
.)
3.
A function block is placed out of service when a Free Place module is selected to fill
the slot occupied by that function block.
0X60
The function block has been placed in simulation mode, using PDM.
0x8E
Value is GOOD, but above HI HI Alarm value in associated Analog Input
block.
0x8A
Value is GOOD, but above HI warning value.
0x8D
Value is GOOD, but below LO LO Alarm value.
0x89
Value is GOOD, but below LO warning value.
0xC4
Bad configuration: value is BAD.
0XDE
AI block out of service: value is BAD.
Floating Point
Status
Distance
Space
Free
Place
—
—
—
—
—
byte 6
byte 7
byte 8
byte 9
byte 10
Free
Place
byte1
byte 2
byte 3
byte 4
byte 5
—
—
—
—
—
SITRANS LR 300.book Page 117 Monday, June 10, 2002 7:18 PM
Summary of Contents for SITRANS LR 300
Page 1: ...SITRANS LR 300 Instruction Manual ITRANS LR 30 July 2002 ...
Page 14: ...Page 12 SITRANS LR 300 INSTRUCTION MANUAL 7ML19985CL02 mmmmm Specifications ...
Page 46: ...Page 44 SITRANS LR 300 INSTRUCTION MANUAL 7ML19985CL02 mmmmm Start Up ...
Page 96: ...Page 94 SITRANS LR 300 INSTRUCTION MANUAL 7ML19985CL02 mmmmm Troubleshooting ...
Page 136: ...Page 134 SITRANS LR 300 INSTRUCTION MANUAL 7ML19985CL02 mmmmm Appendix VIII ...
Page 139: ......