Functions
8.4 Measuring ranges
FIDAMAT 6
92
Operating Instructions, 01/2019, A5E00222135-04
Figure 8-20 Measuring range switchover type A
When the upper switchover point (OU) is exceeded, the next larger measuring range
available is selected. When the lower switchover point (UU) of the next smaller measuring
range available is fallen below, this measuring range is selected. The UU lies at 80 %
(HystA) of the measuring range. The OU lies at 90% (HystE) of the measuring range.
Type B:
The measuring range end value must be larger than or equal to the measuring range end
value following it. Since the measuring spans must also become larger, the measuring range
start values of the following measuring ranges are always smaller.
mba Start of measuring range
mbe End of measuring range
UU
Lower switchover point: Select smaller measuring range
OU
Upper switchover point: Select larger measuring range
Figure 8-21 Measuring range type B
The following applies to the measuring range switchover:
Figure 8-22 Measuring range switchover type B
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......