Functions
8.3 Calibration
FIDAMAT 6
84
Operating Instructions, 01/2019, A5E00222135-04
Note
The settings for "AUTOCAL/-check..." (subfunctions 1 to 3) are valid both for AUTOCAL
(automatic calibration) and for AUTOCAL check (automatic check for adherence to the set
calibration tolerances, without calibration).
The settings for "AUTOCAL Check" (4th subfunction) only refer to the check of the
calibration tolerances without calibration.
AUTOCAL/-check mode
Figure 8-9
AUTOCAL/-check mode
With this subfunction, you configure various AUTOCAL operating modes.
"Autocal/-check on/off":
In
"Autocal off"
state (display:"□"), the settings of
"Start autocal cyclically (parameter)"
and
"Start autocal via binary input"
have no effect on the device.
" Trigger Autocal once"
cannot
be selected. The cycle time continues to run. No automatic calibration is initiated, however.
"Start autocal/-check cyclically":
You can activate AUTOCAL for a regularly recurring cycle if you have set
"Time from Autocal
to Autocal (cycle time)"
using subfunction
"Autocal/-check cyclic parameter"
beforehand.
The AUTOCAL Check is used only for checking the calibrations. As with Autocal, the device
executes the sequence configured in subfunction
"Autocal/Check sequence"
. In contrast to
AUTOCAL, no new calibrations are initiated; only the deviations are checked with respect to
selectable calibration tolerances.
"Start autocal/-check via binary input":
If you have previously configured a binary input with function 72, you can initiate AUTOCAL
via a binary input.
"Trigger Autocal once":
In
"Autocal on"
state, you can start an AUTOCAL sequence at any time with the
"Trigger
Autocal once"
softkey, providing the device is in "Measuring" operating state. A sequence
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......