Functions
8.6 Configuration
FIDAMAT 6
124
Operating Instructions, 01/2019, A5E00222135-04
Interference correction with const. influence of interfering gas
Figure 8-59 Constant influence of interfering gas
With softkey 2, you set whether the cross-interference correction is to apply only to certain
measuring ranges.
Under
"Cross-interference deflection"
, enter the value of the zero offset.
Correction of cross-interference with variable influence of interfering gas
Figure 8-60 Variable influence of interfering gas
Here, a variable influence of interfering gas is active, which must be measured using a
suitable separate gas analyzer and then supplied as an analog current or digital signal (via
ELAN) to the device for correction of cross-interference.
Press softkey 1 again to perform the cross-interference correction via the RS 485 (ELAN)
serial interface.
Here, the same inputs are to be made as for the correction of cross-interference
"…via
analog input"
.
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......