Functions
8.6 Configuration
FIDAMAT 6
128
Operating Instructions, 01/2019, A5E00222135-04
Parame-
ters
Meaning
0
Only the PROFILE Ident number is positively acknowledged
1
Only the device-specific Ident number is positively acknowledged
3
Only the PROFILE Ident number for multivariable devices (complex analyzers) is posi-
tively acknowledged.
Note
In order to work with the provided GSD and EDD, the value 1 must be entered for 'Ident
number'.
●
Relay on PB. This can be used to enable the 8 relays of the additional electronics (option
board) so that they can be remotely controlled via PROFIBUS. For activating, however,
none of these relays are permitted to be assigned a function.
Note
For remote control via PROFIBUS, the software version of the additional PROFIBUS
electronics (option board) must be 2.0.0 or higher.
The current software version of the PROFIBUS card and the boot software version are
indicated in the bottom lines of the display.
8.6.19
Start-up state (function 91)
Figure 8-65 Change start-up state/operating state (FIDAMAT 6-E)
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......