Functions
8.5 Parameters
FIDAMAT 6
104
Operating Instructions, 01/2019, A5E00222135-04
With this function, you clear or lock logbook entries.
From the "Coded/decoded display mode", you can also delete logbook entries with the
button sequence [5][5][5][5] [ENTER].
Status messages such as maintenance requests or faults cannot be suppressed by "Lock
logbook". They still appear, despite the locked logbook.
See also
Logbook (function 3) (Page 78)
8.5.13
Internal valves (function 61)
Figure 8-38 Controlling internal valves
With this function, you switch the internal valves for controlling the gases. Press the
respective softkey to activate ("■") or deactivate ("□") the valves.
The internal valves block or open the gas flow for combustion gas, combustion air, zero gas,
calibration gas and sample gas.
8.5.14
Set external pressure (function 62)
Note
This function is valid only for FIDAMAT 6-G.