Functions
8.6 Configuration
FIDAMAT 6
114
Operating Instructions, 01/2019, A5E00222135-04
4.
Now set the "Span Calib." binary input with the necessary voltage for at least 1 second.
The device calibrates the corresponding measuring range. At the same time, the device
leaves the "Calibrating" state ("□ CAL") and deactivates the "Function control" ("□
CTRL"). It is now back in "Measuring" state. Calibration is complete.
5.
Now set the binary input that you assigned to the desired measuring range after
calibration for at least 1 second.
The device is now in the desired measuring range. You can continue with measurement.
The following figure illustrates the sequence:
Figure 8-46 Calibration of the span via binary inputs
See also
Select measuring ranges (function 40) (Page 90)
Relay outputs (function 71) (Page 107)
Binary inputs (function 72) (Page 109)