Operation
7.6 Calibration
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
73
Zero gas: Synthetic air (free of hydrocarbons)
Step
Procedure
Function
Input
Comment
1
Select start-of-scale and full-
scale values of measuring
range
41
0 – 20 ppm
2
Enter setpoints for zero and
span
22
Zero point: 0 ppm
Span: 15 ppm
Take inlet pres-
sures into account
(Page 167)")
3
Calibrate the zero point
20
4
Calibrating the span
21
b) Detection of hydrocarbons in air
Measuring range: 0 – 50 ppm C3
Calibration gas: 90 ppm methane in nitrogen (6.0)
Zero gas: Nitrogen (free of hydrocarbons)
Step
Procedure
Function
Input
Comment
1
Select start-of-scale and full-
scale values of measuring
range
41
0 – 50 ppm
2
Enter setpoints for zero and
span
22
Zero point:
0 ppm C3 Span:
30 ppm C3
90 ppm C1 yields
30 ppm C3
3
Calibrate the zero point
20
4
Calibrating the span
21
Note
The operator input options of the named functions are described in detail in "Functions".
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......