Functions
8.6 Configuration
FIDAMAT 6
110
Operating Instructions, 01/2019, A5E00222135-04
No binary input has been preassigned in the factory.
Press the fifth softkey ("...cont."), to access the next function screen, thus branching to
further binary inputs.
Note
Save changes
Make sure you save every change in the configuration of the binary inputs in the user data
memory using function 75! If you neglect to do this, there is a danger that a previous
(unwanted) configuration will be loaded by "Load user data" (function 75).
Table 8- 6
Control functions of the binary inputs
Function
Control voltage needed
Remark/effect
0 V 24 V 24 V
(min. 1 s)
Not assigned
No effect when input is set
Fault 1 to 7
N X
e.g. due to:
•
Signal from a gas preparation system:
–
Condensate overflow
–
Gas cooler defective
•
Fault message from interfering gas analyzer
Maintenance
request 1 to 7
N X
Acknowledge
N, X
Setting the input deletes all logbook entries.
However, if the cause of a fault or maintenance re-
quest has not been eliminated, the corresponding
message reappears in the logbook.
Function control
1 to 4
N X
Relay must be configured for function control under
function 71 if, for example, the function is to be con-
trolled by a second device.
AUTOCAL
N, X
Note:
Effective control only possible in "Measuring" operat-
ing state!
Autocal must be configured (function 24)
Measuring range
1 to 4
N, X
When the input is set, the corresponding measuring
range is switched to.
Zero gas
N, X
Note:
Effective control only possible in "Measuring" operat-
ing state!
Relay must be configured for zero gas. After the input
is set, the device changes from "Measuring" to "Cali-
brating" operating state. See also "Example of binary
inputs"
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......