Commissioning
6.3 Commissioning and operation
FIDAMAT 6
58
Operating Instructions, 01/2019, A5E00222135-04
6.3
Commissioning and operation
Procedure
1.
Before you switch on the device, ensure the following:
–
All required gas lines are connected to the device and have been checked for leak
tightness and strength with an overpressure.
–
Combustion gas, combustion air, zero gas and calibration gas are connected and the
required pressures are set.
2.
Switch on the device.
The measured value display appears shortly thereafter on the control panel display.
Above it (in the top line) is the status display (see User interface and operating modes
(Page 65)).
3.
Wait approximately one hour until the warm-up phase of the device is complete.
To prevent blocking the automatic warm-up process, the device must always be in
measuring state. The current operating state can be read in the bottom line.
6.3.1
Warm-up phase
The device requires a warm-up phase of approximately one hour from its cold state.
Note
Do not change any pressures during the warm-up phase.
Warm-up phase
The following sequence only applies if you have not activated another start-up state in
function 91.
Sequence of the warm-up phase
During the warm-up phase, the device runs through the warm-up steps in the following order:
1.
Oven in heating phase
In this phase, the oven warms up (temperature can be called using function 2 "Diagnostic
values").
2.
Chamber in heating phase
The FID is being warmed up.
3.
Combustion gas test
The device checks the combustion gas supply.
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......