Functions
8.6 Configuration
FIDAMAT 6
108
Operating Instructions, 01/2019, A5E00222135-04
Function
Relay is passive when Relay conducts cur-
rent
Note
Measuring range 1 to 4
Measuring range 1 to
4 On
Measuring range identifica-
tion
Limit 1 to 4
Limit 1 (to 4) triggered
Limit alarm on/off
Control function
(CTRL)
Function control active
(e.g. on change to
operator control
mode;
in the warm-up phase,
when AUTOCAL or
calibration is running)
Function control
deactivated
Signal when:
•
Device is decoded
•
Warm-up phase
(max. 2 h)
•
Calibration running
(AUTOCAL)
•
Remote control via inter-
face (ELAN,
PROFIBUS)
Sample gas
Supply of sample gas Actuation of valves
Zero gas 1
Supply of zero gas
Actuation of valves
Calibration gas 1 to 4
Supply of calibration
gas
Actuation of valves
Measuring point 1 to 6
Measuring point 1 to
6 selected
For taking a gas sample via
solenoid valves at various
sample points
Signal for measuring
point 1 to 6
Measuring point 1 to
6 selected
For identifying the sample
point (runs in parallel with
measuring point)
Signaling contact
When signaling, the
relay is briefly ener-
gized.
e.g. in the case of
AUTOCAL:
Control of a second device
Gas path heating
Heating ready for
operation
For identification purposes
Autocal check
Autocal difference
too large (func-
tion 24)
Combustion gas
Combustion gas
valve is open
Actuation of valves
Combustion air
Combustion air valve
is open
Actuation of valves
Operator input prompt
Operator intervention
required
Gas connection faulty;
W5 (combustion
air/combustion gas) pre-
sent;
Device in PAUSE /
STANDBY
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......