Commissioning
6.2 Preparation for commissioning
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
57
Procedure
1.
Connect the gas outlet to a relative pressure manometer (0 to 1000 hPa, resolution
1 hPa).
2.
Generate an overpressure of approximately 750 hPa for about one minute and then stop.
3.
Wait about 10 minutes until the pressure has equalized. Then record the pressure.
4.
Wait another minute, and read off the pressure again.
If the pressure drop within one minute does not exceed 12 hPa, the gas path is sufficiently
tight.
Note
During the measurement, the gas path, including the FID, must have a constant
temperature.
Test conditions:
The specified pressures are relative pressures.
Calibration gas:
Synthetic air (sample gas path must not be contaminated)
Test pressure:
Operating pressure x 1.5 (operating pressure = 500 hPa)
Device electrically switched off:
Test pressure [hPa] relative Pe
rmissible pressure drop Δ p
[hPa] relative
Sample gas path
750 +10
Δp 12
Supply gas inlets
3000 +100
Δp 17
Pressure equalization times: Chronological test sequence
Note
After applying the test pressure, note the start and end of the measurement for determining
Δp.
Start of measurement x min
after
applying test pressure
End of measurement y min after
applying test pressure
Sample gas path
2
3
Supply gas inlets
1
2
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......