5.2
Remote version
Remote installation
MAG 8000 CT user sealing
1. Verify that model and serial numbers shown on labels of sensor and transmitter are
matched properly.
2. Ensure that cable is safely installed to avoid damage of cable and connectors. Please note
the different coil and electrode connector types both having a minimum RADIUS of 45 mm
(1.8 inches); otherwise there s a risk of damaging the cable. Save dust covers for future use
and protection.
3. Ensure connectors are clean.
4. Ensure connectors are fastened securely to achieve a good connection and watertight seal.
Note
If dirt enters connector ends, use plain water for cleaning. Ensure connectors are completely
dry before making connections.
For MAG 8000 CT user sealing, see Sealing of MAG 8000 CT (Page 31).
Connecting
5.2 Remote version
SITRANS FM MAG 8000
Operating Instructions, 03/2020, A5E03071515-AH