MAG 8000 Standard / MAG 8000 CT
MAG 8000 CT (additionally)
Parameter no.
Parameter name
Parameter no.
Parameter name
332
Max. sensor frequency excitation
333
Empty pipe detection
420
Device Communication Address
511
Communication module type
550
Coil current active
551
Fix flow mode active
552
Fixed flow value
Hardware key
To gain access to protected parameters a hardware key must be installed. The hardware key
hole is located in front part of PCB behind battery, see connection diagram (Page 33).
Note
Hardware key on MAG 8000 CT
For MAG 8000 CT only authorized persons, with the acceptance and under direction of the local
authorities are allowed to use the hardware key.
Changing units
Note
Not allowed on MAG 8000 CT
This function is not allowed on MAG 8000 CT.
1. Select service mode and meter version – upload data from meter.
2. Open transmitter.
3. Remove battery (keep connected).
4. Attach hardware key to PCB board (for correct attachment, see upper part of figure).
Flow Tool
C.5 Unit selection
SITRANS FM MAG 8000
128
Operating Instructions, 03/2020, A5E03071515-AH