4.3
Grounding
The sensor body must be grounded using grounding/bonding straps and/or grounding rings to
protect flow signal against stray electrical noise and/or lightning. This ensures that noise is
carried through sensor body and that the measuring area within sensor body is noise-free.
Note
Grounding connector rotated by 90°
For sensor sizes between DN 25 (1") and DN 300 (12"), the grounding connector on the sensor
is rotated by 90°.
Metal pipes
Connect straps to both flanges with 6 mm (1/4") screws.
①
Metal pipes
②
Grounding straps (rotated by 90° on DN 350 to DN 1200)
Grounding straps are part of delivery and pre-mounted on water meter.
Installing/Mounting
4.3 Grounding
SITRANS FM MAG 8000
26
Operating Instructions, 03/2020, A5E03071515-AH