Output B as call-up output
①
Call up output
②
Call up reset
③
Call up status
④
On - Off
When output B is configured as "call-up", the output is activated by an alarm condition and
remains on until it is reset via meter display key or communication interface.
A new alarm will not activate a "call-up" function if the "call-up" function is still active from a
previous alarm.
Note
Like alarm output, call-up output inverts to a pulse output providing a call-up if power
disappears or cable connection is interrupted.
MAG 8000
When output B is configured as an "call-up" output, it will follow the internal alarms that were
previously chosen in the Alarm Configuration List.
Note
Call-up output is inverted to a pulse output providing an alarm if power disappears or cable
connection is interrupted.
Regional factory default settings
DN
mm
size
(inch)
Pulse width
ms
Amount per
pulse
m
3
(Europe)
Amount per
pulse
Gallons (USA)
Amount per
pulse
ML (Australia)
25, 40, 50
(1", 1½", 2")
50
0.01
1
0.001
65, 80, 100, 125, 150
(2½", 3", 4", 5", 6")
50
0.1
10
0.001
200, 250, 300, 350,
400, 450, 500
(8", 10", 12", 14",
16", 18", 20")
50
1
100
0.01
600, 700, 800, 900,
1000, 1050, 1100, 1200
(24", 28", 30", 32", 36",
40", 42", 44", 48")
50
10
100
0.01
Technical data
10.6 Output characteristics
SITRANS FM MAG 8000
100
Operating Instructions, 03/2020, A5E03071515-AH