Maintenance
Make ready for operation:
1
Wait until the sample gas filter has reached operating temperature once more (approx.
15 minutes).
2
Calibrate the device (
14.4.3
Cleaning the filter
Dismantling the filter element:
Fig. 27
Sample gas filter: Dismantling the filter element
1
Remove sealing rings [A] from the face side of the filter element.
2
Remove the screw connection [B]
(on the analyzer unit types “inline” “industrial
housing”:
including the sampling pipe).
3
Remove pressure plate [C].
4
Remove filter element [D].
If the displayed calibration deviations are now unusually high, the sample gas
filter is probably leaking.
●
The filter element is a cylindrical sleeve of sintered metal. Dirt deposits
form on the
inside
of the sleeve.
●
The following applies for the analyzer unit types “inline” and “industrial
housing”: The sampling pipe of the sampling probe is secured directly on
the filter element.
●
Spare parts
CAUTION:
Risk of skin burns
The filter element is hot when the device is in operation (
200 °C).
Before dismantling:
Let the filter element cool down.
C
B
A
D
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...