Installation of the Operating Unit
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
59
6.2
Mains connection of the operating unit
6.2.1
External mains switch
If the mains switch of the operating unit (
) is not visible and accessible in the
operational state:
Install a mains disconnector switch near the operating unit with which the mains supply
can be switched on and off.
Mark this mains switch clearly to prevent confusion.
6.2.2
Mains cable connection
Ensure that the supplied mains voltage corresponds to the information on the type
plate of the operating unit.
Connect the mains cable to the integrated connector on the rear of the operating unit
(
Connect the mains cable to the mains supply.
Fig. 18
Power plug, mains switch
If the operating unit is installed in a 19" system, the mains switch is generally
not visible and cannot be accessed immediately. The European standard
EN 61010 stipulates that an external mains switch must be installed for
stationary devices which do not have an integrated mains switch.
CAUTION:
Damage/malfunction caused by incorrect mains voltage
The operating unit is suitable for a mains voltage of either 230 V or 115 V. If
the mains voltage is too high, the EuroFID can be damaged and constitute a
hazard. If the mains voltage is too low, the function of the EuroFID is impaired.
Check which mains voltage is stated on the type plate.
Only run the EuroFID with this mains voltage.
Technical data for the supply voltage
I / 0
Power
Netz
Fuses
Sicherungen
4A MT
CEE-22
L/N/PE
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...