Project Planning
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
31
3.3
Supply of operational gases
3.3.1
Screw connections of the gas connections
●
Standard:
Clamping ring screw connections for pipes with a 6-mm outside diameter,
brand name “Swagelok”.
●
When the screw connections are removed, inside threads are available:
– Exhaust gas outlet (“Exhaust”): Inside thread G 1/8"
– All other gas connections (except “Bypass”): Inside thread 1/8" NPT
3.3.2
Safety instructions for gas supply
3.3.3
Supply of instrument air
Ensure a continuous supply of instrument air to the “Instr.air” gas connection
(specifications
).
In the event of critical measurement applications (e.g. emission measurement):
Clean
the instrument air (e.g. with a catalytic purifier).
NOTICE:
Contaminated operating gases can corrupt the measured values and damage
the analyzer unit.
Ensure the required purity of all operating gases.
Ensure that the gas pipes to the EuroFID are clean:
– Free from particles (dust, chips)
– Free from hydrocarbons (grease, oil, solvents).
When installing the gas pipes pay special attention to cleanliness
NOTICE:
When the supply of an operating gas fails, the FID flame is extinguished.
Consequence:
– The measuring function of the EuroFID fails.
– Condensation can occur in the internal sample gas paths. As a result the
components of the measuring system (e.g. nozzles) can be rendered
useless.
Ensure a continuous supply of the operating gases (
).
●
Recommendation:
Monitor the supply pressure of the instrument air with
an electronic pressure monitor. Connect the signal of the pressure monitor
with the “Gas supply malfunction” control input (
●
Instrument air is also used as zero gas (
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...